Original Research Article
Development and Validation
of a RP-HPLC Method for Assay of Atorvastatin and its
Application in Dissolution Studies on Thermosensitive
Hydrogel-Based Nanocrystals
Mallesh Kurakula1,
Tariq R Sobahi1, AM El-Helw2 and
Magdy Y Abdelaal1,3*
1Polymer Research Lab,
Department of Chemistry, Faculty of Science, 2Department
of Pharmaceutics, Faculty of Pharmacy, King Abdulaziz
University, Jeddah, Saudi Arabia, 3Department
of Chemistry, Faculty of Science, Mansoura University,
35516-Mansoura, Egypt
*For correspondence:
Email:
myabdelaal@gmail.com,
abdelaal@mans.edu.eg; Tel:
+966-500096707
Received: 26 June 2014
Revised accepted: 8
September 2014
Tropical
Journal of Pharmaceutical Research, October 2014;
13(10): 1681-1687
http://dx.doi.org/10.4314/tjpr.v13i10.16
Abstract
Purpose: To develop and validate a
novel reverse phase high performance liquid
chromatographic (RP-HPLC) method for the quantification
of atorvastatin in thermosensitive hydrogel-based
nanocrystal formulation.
Method: Chromatographic
identification was achieved on C18 (5 µm) column using
acetonitrile and 0.025 M potassium dihydrogen ortho-phosphate
buffer pH 5 (45:55 v/v) as mobile phase, at a flow rate
of 1.5 mL/min and using photo diode array detector (PDA)
at 246 nm. The developed HPLC method was validated
according to International Conference on Harmonisation (ICH)
Q2(R1) guidelines and applied to dissolution studies on
atorvastatin thermosensitive hydrogel-based nanocrystal
formulation, using Lipitor® as standard.
Results: Determination was
successfully achieved with good peak resolution from
atorvastatin nanocrystals and a commercial formulation
brand (Lipitor® tablets) without interference of polymer
or excipients. The retention time of atorvastatin was
4.5 min and drug response was linear in the range of 0.1
- 0.5 µg/mL with a correlation coefficient of 0.9995.
Precision was determined to be between 0.16 - 0.61
percent relative standard deviation (% RSD) for the
analyzed samples. The limit of detection and of
quantification was 35.6 and 71.2 ng/mL, respectively,
which was 10 times higher than a previously reported
method. The assay of atorvastatin nanocrystal and
Lipitor® gave 99.37 and 99.12 % recovery, respectively.
Dissolution studies showed atorvastatin release of 40
and 65 % at 40 min from thermosensitive hydrogel
nanocrystal formulation and Lipitor®, respectively
indicating sustained release.
Conclusion: The method is successfully
validated and is specific, linear, precise, and accurate
with good robustness. It is applicable to atorvastatin
nanocrystal dissolution studies and is a promising
quality control tool for atorvastatin analysis in
nanoformulations and pharmaceutical dosage forms.
Keywords: Atorvastatin,
Anticholestermic, Dissolution studies, Hydrogel,
Nanocrystal, Thermosensitive |